Maintenance Pages


NMR Labs Latest Calibration Values

 

NMR Instrument

INOVA500

MERCBB

MERCURY

VXR300

gemini300

TBA

installed  probe

atb

asw

zspec

5mmSW

Lineshape test

.19,2.86,4.92

.26, 2.65, 4.84

.63,3.09,5.09

.27,2.07,3.51

Nucleus

SN@pw90/tpwr

SN@pw90/tpwr

SN@pw90/tpwr

SN@pw90/tpwr

Proton

171@9.9/59

116@12.6/59

43.4@15/59

108.5 @ 10.0/55dB

Carbon

70.5@6.8/63

83.4@14.75/59

54.75@4/55

85.9@10.12uS/59dB

Flourine

148@18/59

78.4@25.25/62

45.1@18.75/59

NA

Phosphorus

NA-Self-Xtune

82.5@26/59

49.0@4/55

NA-Self-Xtune

NMR User Notes

Excellent Shims

Excellent SN

Excellent Shims

Very Good 4nuc SN

Very Good Shims

Fair 4nuc SN

Excellent Shims

Very Good H/C SN

 

Line-shape test uses latest shims and measures width in Hz of the NMR singlet in 20% CdCl3/Acetone @ 50%, .55%, .11% line-height respectively.  Lower is better and implies better signal separation.

 

Signal to Noise (SN) comparisons with higher values are better (ignore pw90/tpwr).  For long duration experiments lower values add significant time.  For a certain signal to noise value found using command dsn (display signal to noise) you may expect ~ a doubling of signal by quadrupling the nt (number of transients).  Any questions on interpreting the above table please email, call or stop by.

 

 

Robert Harker

Engineer

UF CLAS – Chemistry NMR Labs

 Office:  CLB106 => Hrs 8:30 - 5PM Most workdays

Gainesville Fl 32611-7200


Page created and maintained by Robert Harker.

Last update: Novemeber 29, 2007 Robert Harker